Description

Ando 6315E Optical Spectrum Analyzer (OSA)

OSA has passed all manufacturer specifications except for Dynamic Range for Single Pass Monochromator. Calibration report available upon request.

Calibration specs results

  • Wavelength Accuracy: passed
  • Wavelength Linearity: passed
  • Wavelength Reproducibility: passed
  • Level Accuracy: passed
  • Level Linearity: passed
  • Polarization Dependency: passed
  • Measurement Level Range: passed
  • Wide Band Level Stability: passed
  • Dynamic Range Double Pass Monochromator: passed
  • Dynamic Range Single Pass Monochromator: failed

Specifications at a glance for the 6315E

  • Wavelength Range: 350-1750 nm
  • Dynamic Range: 70 dB
  • Wavelength Accuracy: ± 0.5 nm (350-1750 nm)
  • Wavelength Linearity: ± 0.02 nm (1510 to 1570 nm)
  • Wavelength Repeatability: ± 0.005 nm
  • Polarization Dependency: ± 0.01dB

Additional Information

  • Wavelength Resolution: 0.05 to 10 nm
  • Level Accuracy: ± 0.3 dB
  • Level Linearity: ± 0.05 dB
  • Sweep Time: 2.5 sec or less (350-1750 nm)