You may also like

Ando (Yokogawa) AQ6315A cust 001 Optical Spectrum Analyzer(OSA)

The Ando (Yokogawa) AQ6315A Optical Spectrum Analyzer (OSA) is a device that brings advanced capabilities designed for precise optical analysis across a wide wavelength range (350-1750 nm). Featuring significant upgrades, Ando incorporate a totally new monochromator design. This device delivers exceptional wavelength accuracy (±0.05 nm) and dynamic range. Ideal for WDM, LED, and fiber optic applications, this analyzer is perfect for professionals requiring reliable, high-sensitivity measurements with advanced functionality. Its user-friendly interface and diverse analysis tools make it a trusted solution for optical evaluation.

Download Ando AQ6315A SpecificationsDownload Ando AQ6315A Specifications

Download Ando AQ6315A User ManualDownload Ando AQ6315A User Manual

Read moreShow less

Request Pricing

5 1
,
5/ 5
Out-of-Stock
Condition
Used
Reference:
AQ6315A_CUST001
Brand:
Description
Ando (Yokogawa) AQ6315A Optical Spectrum Analyzer (OSA) - Precision Meets Innovation

The Ando (Yokogawa) AQ6315A Optical Spectrum Analyzer (OSA) is a compact OSA, it supports signle-pass/double-pass monochrometer mode select. In signle-pass monochrometer mode, it supports the same wide range of applications as the predecessors. In double-pass monochrometer mode, dynamic range covers from 350nm to 1750nm. Built for professionals in optical research, telecommunications, and component testing, this analyzer is engineered to deliver good accuracy and versatile performance. Whether you're evaluating light sources, assessing wavelength characteristics, or testing WDM devices, the AQ6315A is designed to exceed your expectations.

Calibration of the Ando (Yokogawa) AQ6315A Optical Spectrum Analyzer (OSA)
  • The Ando (Yokogawa) AQ6315A Optical Spectrum Analyzer (OSA) is calibrated to OEM specifications prior to shipping.
  • Each unit includes a Calibration Certificate, ensuring traceable and reliable calibration results.
  • A Calibration Label is affixed to the device for quick verification of calibration status.
  • Detailed Calibration Data is provided to support accurate and precise measurements.
Specifications of the Ando (Yokogawa) AQ6315A Optical Spectrum Analyzer (OSA)
Feature Details
Wavelength Range 350 nm to 1750 nm
Wavelength Accuracy ±0.05 nm (1510 to 1570 nm)
Wavelength Linearity ±0.02 nm (1510 to 1570 nm)
Dynamic Range Single-pass monochromator mode:
40dB (±1 nm, 633/1152/1523 nm)
35dB (±0.5 nm, 633/1152/1523 nm)
Double-pass monochromator mode:
70dB (±1 nm, 633/1152/1523 nm)
60dB (±0.5 nm, 633/1152/1523 nm)
Wavelength Repeatability ±0.005 nm
Polarization Dependency Single-pass monochromator mode:
±0.01 dB (1310 to 1550 nm)
Double-pass monochromator mode:
±0.05 dB (1310 to 1550 nm)
Wavelength Resolution 0.05 to 10 nm
Level Accuracy ±0.3dB (at 633/1310/1550 nm)
Level Linearity ±0.05dB (0 to -40dBm)
Sweep Time 0.5 s or less (span = 500 nm), 2.5 s or less (full span)
Max Input Power +20 dBm (100 mW)
Calibration Calibrated to OEM specifications, includes Certificate, Label, and Data

With its comprehensive features and user-oriented design, the Ando (Yokogawa) AQ6315A Optical Spectrum Analyzer (OSA) is the perfect solution for professionals seeking precision, versatility, and reliability in optical spectrum analysis.

Read moreShow less
Product Details
AQ6315A_CUST001
Terms and Conditions

Settings